BTEXs in water samples were determined with VF-5 ms and CP-8ms capillary-column and Headspace-Gas Chromatography/Mass spectrometer (HS-GC/MS) under modified conditions, especially with the headspace. Results show that when 10 ml flasks with water sample accounting for 60% of each flask in volume were agitated for 20 sec. at a rate of 500 r min -1, in the heated shaker, then kept still for 90 sec. mixed with 2.4 g NaCl in each flask and left for equilibrium for 30 min. under 60 ℃, optimal results could be obtained. The BTEXs detection limits of the method varied the range of 0.10~0.13 μg L -1; the standard curve correlation coefficient (r) of each BTEXs target was more than or equal to 0.999; and the sample recovery rate lingered in the rang of 98.4%~100.2% with precision being 1.41%~6.52% (n=7). All demonstrate that the method meets the requirements of organic analysis.
Reference
Related
Cited by
Get Citation
Chen Hong, Han Yong, Wang Xi, Zhong Ming, Jin Wei. Optimized determination of BTEXs in water samples with headspace-gas chromatography/mass spectrometer[J]. Acta Pedologica Sinica,2013,50(5):992-997.